Measuring principle
In voltammetry the potential is set and the current is measured. In EFCE the current is additionally controlled by an external field; the controller output is measured.
- Working electrode
- Pt microdisc, Ø 5 µm, embedded
- Electrolyte
- 0.05 M H₂SO₄
- Field electrode
- Pt wire, Ø 10 µm, behind 6 µm muscovite mica
- Reference
- Hg/Hg₂SO₄, chloride-free; two-electrode operation
- Control
- I controller, 100–1000 Hz, FPGA-based; manipulated variable field, controlled variable faradaic current
- Conditioning. Slow potential cycles largely remove the outer double layer; the inner Helmholtz layer is retained.
- Potential sweep. The potential is varied slowly while the controller holds the faradaic current at the setpoint.
- Capture threshold. Potential from which the control first takes effect at low field strength.
- Control curve. Field strength as a function of potential above the capture threshold.
The control curve is independent of the setpoint current: anodic and cathodic setpoints as well as I = 0 yield the same curve. We therefore interpret it as a representation of the interfacial state rather than of the faradaic turnover.
First description: Schönweitz & Schönweitz, “Electric Field Coupled Electrode”, Electroanalysis 11 (1999) 957–963.